3
Pegasus firmware was updated to record VEI in millivolts.
# Firmware notices
Previous versions of the firmware recorded VEI in microvolts.
We decided to apply VEI's conversion factor in newer version of Pegasus.
---------------------------
If user observe values higher than expected values for Pegasus, the data is of the older version.
---------------------------
\ No newline at end of file
 
## Old Pegasus firmwares
 
 
Some old Pegasus firmwares record a number of SOH channels in different units.
 
These channels included:
 
+ VEI: microvolts in old firmware, millivolts in new firmware
 
+ VDT: microdegrees C in old firmware, centidegrees C in new firmware
 
+ VM1-6: microvolts in old firmware, millivolts in new firmware
 
 
Currently, SOHViewer only supports new versions of the Pegasus firmware. If the
 
plotted values of the channels listed above are higher than expected, the data
 
is recorded using old firmwares. This problem can be fixed by editing the
 
affected channels' conversion factor to reflect the old units (refer to the
 
View-Edit Database Tables section).